This document specifies the requirements for washer-disinfectors (WD) intended to be used for the cleaning and chemical disinfection, in a single operating cycle, of reusable items such as:
u000b
a) bed frames;
u000b
b) bedside tables;
u000b
c) transport carts;
u000b
d) containers;
u000b
e) surgical tables;
u000b
f) sterilization containers;
u000b
g) surgical clogs;
u000b
h) wheelchairs;
u000b
i) aids for persons with disabilities.
u000b
This document also specifies the performance requirements for the cleaning and disinfection of the WD and its components and accessories.
u000b
Devices identified within the scopes of ISO15883-2, ISO15883-3, ISO15883-4, and ISO15883-6 do not fall within the scope of this document.
u000b
In addition, this document specifies the methods for type testing, works testing, validation (installation, operation, and performance qualification on first installation), routine control, and monitoring, as well as requalifications to be carried out periodically and after essential repairs.
u000b
NOTE 1 WD covered by this document can also be used for cleaning and chemical disinfection of other thermolabile and reusable devices as recommended in the instructions for use (IFU) for those devices.
u000b
NOTE 2 The performance requirements specified in this document cannot ensure the inactivation or removal of the causative agent(s) (prion proteins) of transmissible spongiform encephalopathies.